ALARMS
Once an alarm is activated at the transmitter it will display which input is in alarm and the sounder will be activated. The AS200RX receiver will display both the unit number and input of the transmitter alarming; these details will be time and date stamped. Any communications equipment attached to the receiver will be activated immediately or after a programmable window of 10 minutes.
The system has a reactivation circuit which will re-alarm at the receiver after 1 hour if the transmitter in alarm has not been muted. This stops alarms being ignored as the user must mute the transmitter at the equipment end.  The reactivation facility can be disabled or enabled at the receiver.
 
TEMPERATURE RANGE
The temperature range of the transmitter is –200°C to +100°C with an accuracy of 0.1°C across the entire range.  This makes the transmitter suitable for many types of equipment from LN2 to incubators.
 
FRONT PANEL DISPLAY
The front panel display can be used to show the unit number of the transmitter or time delay or temperature of input 1.  When in an alarm condition the display toggles between alarm I/P and variable selected.  The status led indicates the current condition of the transmitter and the alarm active led flashes brightly in an alarm condition.
 
HIGH & LOW ALARMS
A single high and low alarm can be set for I/P1 and I/P2, these parameters can easily be displayed.
 
DELAY TIME
A single delay time can be set for I/P 1 and I/P 2 ranging from 0 to 90 minutes, these parameters can easily be displayed.  Input 3 has a fixed delay time of 10 minutes.
 
INPUTS
Up to three inputs are available on the transmitter.  A common configuration would be to use input 1 to monitor the air temperature of the equipment, input 2 to monitor the sample temperature and input 3 to monitor the door open / closed condition.  This configuration of the inputs will give maximum protection and temperature data for the sample will be recorded as well as the air temperature if logging is employed on the system.
Input 1 : This is for temperature and the value can be displayed on the transmitter.  This input sends its temperature value to the receiver so it can be monitored and logged if required.
Input 2 : This is for temperature.  This input sends its temperature value to the receiver so it can be monitored and logged if required. (This input is an option)
Input 3 : This is a volt free contact, normally closed in the OK condition and normally open in the alarm condition.
 
POWER FAILURE ALARM
If power failure is detected for longer than 10 minutes the transmitter will activate a power failure alarm at the transmitter and receiver.
 
CALIBRATION
The temperature transmitter has been designed to have an easy calibration procedure.  The transmitter display will show a calibration reminder automatically which is customer enabled, the reminder can be set for intervals of 6 or 12 months or be disabled.  Traceable calibration for temperature can be carried out for the AS200TX alarm transmitters.
 
AS200TS ► TEST TEMPERATURE FOB
Test temperatures at +50.0°C, 0.0°C and  -50.0°C can be simulated on the transmitter input 1 to test the alarm.  The fob temperature simulator is attached to a connector inside the transmitter unit and a temperature probe bypass switch is thrown.  The simulated temperature is selected and will be shown on the display.  If this temperature is out with the alarm limits it will activate the relevant alarm after the time delay.
 
DATA LOGGING & MONITORING
Data logging and monitoring from AS200TX transmitters is possible with the AS200RX receiver when used in conjunction with a computer and relevant software.
The data collected for each transmitter is as follows :-  Serial Number, Unit Number, Temperature I/P1, Temperature I/P2, High Alarm, Low Alarm and Delay Time.
 
The transmitter will update the computer every 30 minutes with the data. This increases to every 10 minutes when the transmitter is in an alarm condition. Faster data collection
n available on request.
 
Data capture software enables the display of variables collected.  A range of software is available to cover clients requirements including :-

· Trend Graphs

· Multi-graphs

· Printed Reports

· Recipe & Batch Facility

· SMS & Email Facilities

· FDA 21 CFR 11 Compliance

· Electronic Records and Signatures